LSR II troubleshooting
- Neither cytometer nor computer will not turn on.
- Make sure that UPS under bench is turned on.
- Can't log in to computer.
- Make sure you are logging on to the CCMI domain.
- Make sure you put in the correct user name and password.
- Computer will not connect to cytometer.
- Restart Diva software.
- Computer STILL will not connect to cytometer.
- Restart Diva hardware: Open the "Cytometer Status" application (the red telephone on the Windows
desktop) and press the Enter key a few times. If you get a prompt such as "-->", type
"reboot" and press Enter. If this fails, cycle the power to the cytometer.
- No PE signal.
- Ensure the 561 nm laser is turned on.
- No events.
- Laser(s) not turned on.
- Fluidics control not set to "RUN".
- Tube is cracked.
- Check level in sheath tank.
- Ensure that sheath tank lid is fitted properly.
- Stuck fluidics - PRIME.
- No cells in sample.
- Clumpy sample - filter it.
- Blocked SIP or flow cell - Remove sample, and press PRIME.
- Waste not venting - check waste line for flow.
- Air bubbles in sheath filter - purge filter.
- Sample runs for a few seconds then stops
- Clumpy sample - filter it.
- Blocked SIP or flow cell - PRIME.
- Waste not venting - check waste line.
- Air bubbles in sheath filter - purge filter.
- Very slow sample rate.
- Sample very dilute.
- Fluidics set to LOW.
- Fluidics fine control turned down.
- Sample still will not run
- Air bubbles in sheath filter - purge filter.
- SIP blocked - Set fluidics to "HIGH". Run 100% bleach for 30 sec. followed by at least 2 changes of DI water for 1 minute.
|