Facscan troubleshooting

Neither cytometer nor computer will not turn on.
Make sure that UPS under bench is turned on.
Computer will not connect to cytometer.
Restart computer.
Cytometer status is "NOT READY".
Allow to warm up for 5 minutes.
Check sheath is full and waste is empty.
If status window says that waste is full and you know it is empty, rap on waste tank to dislodge sensor float.
Check that tubing leading to tanks is not pinched or obstructed.
Cytometer status remains at "STANDBY".
Fluidics control not turned to "RUN"
Tube is cracked.
Check that pressure is ON (vent valve UP).
Check that sheath tank is actually pressurized - it should not be moveable.
Ensure that sheath tank cap is screwed down tightly.
Stuck fluidics - turn to STANDBY and back to RUN.
Status READY, but sample will not run.
No cells in sample.
Clumpy sample - filter it.
Blocked SIP or flow cell - Remove sample, then DRAIN/FILL 3 times.
Waste not venting - loosen waste tank cap. Report problem.
Sample runs for a few seconds then stops.
Clumpy sample - filter it.
Blocked SIP or flow cell - Remove sample, then DRAIN/FILL 3 times.
Waste not venting - loosen waste tank cap. Report problem.
Sample still will not run.
Air bubbles in sheath filter - purge filter.
SIP blocked - Set fluidics to "HIGH". Run 100% bleach for 30 sec. followed by at least 2 changes of DI water for 1 minute.